Instructions for Use Fabius MRI SW 3.n
213
Technical Data
S-ORC (Sensitive Oxygen Ratio Controller)
S-ORC is a control element which guarantees a minimum O
2
concentration in the fresh gas flow. As
from a flow rate of approx. 300 mL/min, the N
2
O concentration in the fresh gas can be freely set
between 0 and 75 %.
During O
2
shortage
S-ORC limits the N
2
O concentration in the fresh gas, so that the O
2
con-
centration does not drop below 23 vol.%.
N
2
O metering valve open
and O
2
metering valve
closed or O
2
flow less than
0.2 L/min
S-ORC prevents N
2
O flow
During N
2
O failure
O
2
may still be administered. No alarm.
Serial Interface
Type: RS - 232
Baud Rates: 1200, 2400, 4800, 9600, 19.2, 38.4 K
Parity: Odd, Even, None
Data Bits: 7 or 8
Stop Bits: 1 or 2
Protocol: MEDIBUS
Summary of Contents for Fabius
Page 36: ...36 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 84: ...84 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 130: ...130 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 192: ...192 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 215: ...Instructions for Use Fabius MRI SW 3 n 215 Technical Data 311 ...
Page 216: ...216 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 218: ...218 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 226: ...226 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 228: ...228 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 231: ...Instructions for Use Fabius MRI SW 3 n 231 This page intentionally left blank ...