Instructions for Use Fabius MRI SW 3.n
77
Assembly
Positioning
z
Place the Fabius MRI at the desired position
(right or left side). Teslameter sensorics, located
at the middle of both sides, will generate an
acoustic alarm when the 40 mtesla fringe field
strength is exceeded. The Fabius MRI must be
moved away from the MRI system until the
acoustic alarm is silenced.
z
Indicate the positon of the device on the floor
using hazard warning tape (black/yellow).
Following the setup and positioning of the
Fabius MRI, perform the imaging test protocol pro-
vided in Appendix 2 of this manual.
WARNING
After positioning, lock the castors on the trol-
ley by central brake at all times in the MRI
room. Magnetic attractive forces between the
magnet and the anesthesia machine may
cause unintentional movement of the anesthe-
sia machine if the casters are unlocked.
NOTE
The two teslameter sensorics (
1
in illustration)
detect and alarm independently.
CAUTION
Make sure that the additional alarm lights at the
edges of the top shelf of the Fabius MRI are
always visible.
CAUTION
Perform the imaging test protocol every time after:
– Fabius MRI is set up for the first time
– additional parts or any device is mounted on
the Fabius MRI for the first time
– maintenance is performed
This ensures that the Fabius MRI does not inter-
fere with the MRT system.
See the test protocol provided in Appendix 2 of
this manual.
102
1
1
Summary of Contents for Fabius
Page 36: ...36 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 84: ...84 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 130: ...130 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 192: ...192 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 215: ...Instructions for Use Fabius MRI SW 3 n 215 Technical Data 311 ...
Page 216: ...216 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 218: ...218 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 226: ...226 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 228: ...228 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 231: ...Instructions for Use Fabius MRI SW 3 n 231 This page intentionally left blank ...