Operation
CLIC Adapter (Optional)
The disposable CLIC adapter absorber can also be
used on the Fabius MRI. For information on install-
ing the CLIC adapter, consult its Instructions for
Use.
Points to note when using the CLIC adapter
with neonates and infants
The compliance of the breathing system is reduced
when the disposable absorber is removed. As a
result, the ventilator delivers a higher tidal volume
to the patient during IPPV/SIMV and the airway
pressure increases accordingly. The effect on the
tidal volume and airway pressure is negligible in
adults with a lung compliance of approx.
50 mL/cmH
2
O. In neonates and infants with a lung
compliance of approx. 5 mL/cmH
2
O or less, the
tidal volume and airway pressure can increase con-
siderably and harm the patient if the disposable
absorber is removed during IPPV/SIMV.
Before removing the disposable absorber:
z
Set
P
MAX
to the plateau pressure.
As soon as the absorber has been
replaced:
z
Set
P
MAX
to the original value again.
Summary of Contents for Fabius
Page 36: ...36 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 84: ...84 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 130: ...130 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 192: ...192 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 215: ...Instructions for Use Fabius MRI SW 3 n 215 Technical Data 311 ...
Page 216: ...216 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 218: ...218 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 226: ...226 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 228: ...228 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 231: ...Instructions for Use Fabius MRI SW 3 n 231 This page intentionally left blank ...