Configuration
144
Instructions for Use Fabius MRI SW 3.n
The adjustment range and factory default values for
all alarms on the Fabius MRI are shown in the fol-
lowing table.
Alarm Volume
z
Select and confirm the »
Alarm Volume
« label
on the Standby Setup screen. The current alarm
volume value appears on the screen.
z
Select and confirm the new alarm volume value
from 1 (minimum) to 10 (maximum)
(range: >45 dB(A) to <85 dB(A)).
In the example illustration, the value was changed
from »
4
« to »
5
«.
The Alarm Volume Setting window is then removed
from the screen, and the cursor returns to the return
arrow.
Alarm
Parameter
Adjustment
Range
Factory
Default Value
O
2
%
19 to 100
18 to 99
100
20
MV
L/min
0.1 to 20.0
0.0 to 19.9
12.0
3.0
Pressure
cmH
2
O
(hPa)
10 to 70
5 to 30
40
8
540
539
Summary of Contents for Fabius
Page 36: ...36 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 84: ...84 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 130: ...130 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 192: ...192 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 215: ...Instructions for Use Fabius MRI SW 3 n 215 Technical Data 311 ...
Page 216: ...216 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 218: ...218 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 226: ...226 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 228: ...228 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 231: ...Instructions for Use Fabius MRI SW 3 n 231 This page intentionally left blank ...