Cleaning
178
Instructions for Use Fabius MRI SW 3.n
Reassembling the Breathing System
Attaching the Inspiratory Valve
1
Place the gasket on top of the valve disc.
2
Place the valve disc in the valve seat.
3
Fit the inspection cap (with port).
4
Tighten the retaining nut securely.
5
Insert the plug for the inspiration dome.
Attaching the Expiratory Valve
6
Place the gasket on top of the valve disc.
7
Place the valve disc in the valve seat.
8
Fit the inspection cap.
9
Tighten the retaining nut securely.
Attaching the APL Valve
10
Position the APL valve in the valve seat, and
tighten securely with the retaining nut.
WARNING
Route all lines/cables away from the APL
valve to prevent interference with the APL
valve adjustment knob. Lines/cables caught
underneath the APL valve adjustment knob
could interfere with proper functioning of this
valve.
060
1
2
3
4
6
7
8
9
5
046
10
Summary of Contents for Fabius
Page 36: ...36 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 84: ...84 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 130: ...130 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 192: ...192 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 215: ...Instructions for Use Fabius MRI SW 3 n 215 Technical Data 311 ...
Page 216: ...216 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 218: ...218 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 226: ...226 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 228: ...228 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 231: ...Instructions for Use Fabius MRI SW 3 n 231 This page intentionally left blank ...