Instructions for Use Fabius MRI SW 3.n
29
System Overview
Teslameter
To ensure that the Fabius MRI works safely and
without interference in the proximity of an MRT sys-
tem, and that the MRT’s imaging is not influenced
out of margin, there is defined distance that the
Fabius MRI machine must maintain from the MRT
system. For exposition in a homogenous magnetic
field, there is a limit of 40 mtesla (400 gauss) to be
kept for the Fabius MRI.
The Fabius MRI is a floor-based workstation that
can be moved around the room as required by the
patient’s respiration/anesthesia during the MRT
treatment. If the fringe field strenght is exceeded,
an acoustic signal is annunciated and continues to
annunciate until the Fabius MRI is repositioned to
the required limit of 40 mtesla (400 gauss). The
array for detecting the magnetic field is a position-
ing aid towards the MRT magnet.
Location of the Sensors
There are two sensors (one on each side) located
in the lower part of basic unit (
1)
.
Checking the sensors in self-test mode, e.g., after
switching on the machine, is not necessary. Both
Tesla sensors are constantly active when the
Fabius MRI main power switch is turned on, includ-
ing in Standby mode and operation on battery
power.
102
1
1
Summary of Contents for Fabius
Page 36: ...36 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 84: ...84 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 130: ...130 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 192: ...192 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 215: ...Instructions for Use Fabius MRI SW 3 n 215 Technical Data 311 ...
Page 216: ...216 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 218: ...218 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 226: ...226 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 228: ...228 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 231: ...Instructions for Use Fabius MRI SW 3 n 231 This page intentionally left blank ...