Operation
102
Instructions for Use Fabius MRI SW 3.n
Pressure Support Ventilation
Pressure Support ventilation is intended to reduce
the work of breathing and is indicated for use only
in patients who are breathing spontaneously.
Patients who are not making spontaneous breath-
ing efforts are not candidates for Pressure Support
ventilation.
Pressure Support ventilation is triggered by the
patient's spontaneous effort to breathe. Most anes-
thetic agents will cause patients to have reduced
ventilatory responses to carbon dioxide and to
hypoxemia. Therefore, patient triggered modes of
ventilation may not produce adequate ventilation.
Additionally, the use of neuromuscular blocking
agents will interfere with patient triggering.
Apnea Ventilation is a feature within Pressure Sup-
port ventilation. To enable Apnea Ventilation, adjust
the
Freq Min
setting to a value other than
OFF
. If
the detected patient spontaneous breathing rate
falls below the set value, the ventilator automati-
cally delivers a Pressure Support breath.
Apnea ventilation is intended to provide some
degree of gas exchange if the patient’s respiratory
rate falls below the desired minimum setting. It is
not intended as a primary mode of ventilation.
When delivering Apnea Ventilation, the Fabius MRI
uses the Pressure Support settings for
Δ
P
PS
,
Freq Min
,
Insp Flow
, and
PEEP
.
If two consecutive Apnea Ventilation breaths occur,
the Caution message »
APNEA VENTILATION!!
«
appears in the Alarm window. The alarm is cleared
when a spontaneous breath is detected.
Summary of Contents for Fabius
Page 36: ...36 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 84: ...84 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 130: ...130 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 192: ...192 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 215: ...Instructions for Use Fabius MRI SW 3 n 215 Technical Data 311 ...
Page 216: ...216 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 218: ...218 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 226: ...226 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 228: ...228 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 231: ...Instructions for Use Fabius MRI SW 3 n 231 This page intentionally left blank ...