Configuration
152
Instructions for Use Fabius MRI SW 3.n
Configuration during Operation
The user can perform O
2
calibrations and view and
change certain monitoring settings for the current
operation while in Volume Control, Pressure Con-
trol, Pressure Support, SIMV/PS, and Man/Spont
mode.
1
Press the »
Setup
« key while in Volume Control,
Pressure Control, Pressure Support, SIMV/PS,
or Man/Spont mode.
z
The Setup window replaces the waveform and
soft keys for the current ventilator mode.
The following soft key labels appear in the Setup
window:
– »
Volume Alarms On/Off
«
– »
Auto Set
«
– »
Calibrate O2 Sensor
«
– »
Des Comp On/Off
«
– »
Access Alarm Log
«
– »
Access Alarm Volume
«
There is a 15-second timeout period for any of the
Setup functions during operation. If no rotary knob
activity occurs within 15 seconds, the Setup win-
dow is removed and the waveform window is redis-
played. The waveform window can be also be
displayed by pressing the »
Home
«
key.
NOTE
To set default monitoring settings to be used at the
power-up of each operation, see “Standby Setup
Screen” on page 140.
566
1
503
Summary of Contents for Fabius
Page 36: ...36 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 84: ...84 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 130: ...130 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 192: ...192 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 215: ...Instructions for Use Fabius MRI SW 3 n 215 Technical Data 311 ...
Page 216: ...216 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 218: ...218 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 226: ...226 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 228: ...228 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 231: ...Instructions for Use Fabius MRI SW 3 n 231 This page intentionally left blank ...