Instructions for Use Fabius MRI SW 3.n
89
Operation
O
2
Flush
A manually operated O
2
flush valve is located on
the front of the machine. When actuated, the valve
delivers an unmetered flow of at least 35 L/min to
the breathing system and breathing bag while
bypassing the ventilator. The Fabius MRI does not
need to be switched on in order to use the O
2
flush.
3
Press the O
2
flush button. Additional O
2
flows
into the compact breathing system as long as
the button is pushed in. The flow control ele-
ments and the anesthetic agent vaporizer are
bypassed.
NOTE
A NOTE provides additional information intended
to avoid inconvenience during operation.
037
3