Fault-Cause-Remedy
162
Instructions for Use Fabius MRI SW 3.n
Fault-Cause-Remedy
The Fabius MRI divides alarm messages into three
categories based on priority. Exclamation marks
indicate the priority. The alarm messages are dis-
played on colored background if the option "Color
display" is enabled:
Internal priority numbers, see page 119, for the
ranking within an alarm category are written below
in brackets, e.g. (23 / 31).
!!!
Warning (red)
high priority
!!
Caution (yellow)
medium priority
!
Advisory (white)
low priority
Priority Alarm Message
Probable Cause
Remedy
!!!
(31)
AIRWAY PRESSURE HIGH
Upper alarm limit for airway
pressure has been exceeded,
ventilation hose is kinked.
Alarm limit has been set too
low.
Check hose system on anes-
thesia machine.
Check breathing circuit or
alarm limit value.
!! / !!!
(23/31)
APNEA FLOW
1)
1) APNEA FLOW alarm priorities are based on alarm duration and ventilation mode:
In Volume Control, Pressure Control, SIMV/PS with Freq
≥
6, or Pressure Support with Apnea Ventilation OFF:
Caution = V
T
<20 mL for >15 seconds
Warning = V
T
< 20 mL for >30 seconds
In ManSpont, SIMV/PS with Freq <6, or Pressure Support with Apnea Ventilation ON:
Caution = V
T
<20 mL for >30 seconds
Warning = V
T
<20 mL for >60 seconds
Breathing/ventilation stops.
Breathing circuit disconnected
or leakage.
Check ventilator.
Check breathing circuit.
!! / !!!
(23/31)
APNEA PRESSURE
2)
2) APNEA PRESSURE alarm priorities are based on alarm duration and ventilation mode:
In Volume Control, Pressure Control, SIMV/PS with Freq
≥
6, or Pressure Support with Apnea Ventilation OFF:
Caution = P
AW
does not cross the pressure threshold for >15 seconds
Warning = P
AW
does not cross the pressure threshold for >30 seconds
In ManSpont, SIMV/PS with Freq <6, or Pressure Support with Apnea Ventilation ON:
Caution = P
AW
does not cross the pressure threshold for >30 seconds
Warning = P
AW
does not cross the pressure threshold for >60 seconds
Breathing/ventilation stops.
Breathing circuit disconnected
or leakage.
Check ventilator.
Check breathing circuit.
Summary of Contents for Fabius
Page 36: ...36 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 84: ...84 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 130: ...130 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 192: ...192 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 215: ...Instructions for Use Fabius MRI SW 3 n 215 Technical Data 311 ...
Page 216: ...216 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 218: ...218 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 226: ...226 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 228: ...228 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 231: ...Instructions for Use Fabius MRI SW 3 n 231 This page intentionally left blank ...