Instructions for Use Fabius MRI SW 3.n
159
Fault-Cause-Remedy
When the battery is completely discharged
Fabius MRI switches off. As a consequence all indi-
vidual settings, which are not saved in the default
settings, will be lost.
All pneumatic functions of the Fabius MRI continue
to be available (APL valve, breathing pressure
gauge, cylinder and pipeline gauges, fresh gas and
agent delivery, S-ORC, and total flowmeter). Man-
ual or spontaneous ventilation can be maintained.
When the power supply is recovered and
Fabius MRI restarted, all ventilation and alarm set-
tings are reset to the saved default settings.
WARNING
When the »BATTERY LOW!!« Caution alarm
message is first displayed, the ventilator will
continue to operate for up to an additional
10 minutes. Then, automatic ventilation is not
available until AC power is restored.
CAUTION
Never allow the battery to completely discharge. If
the battery is discharged completely, recharge
immediately. Do not use the device until the bat-
tery is recharged completely as injury to the
patient may occur if the battery is unavailable.
WARNING
Danger of patient injury by wrong device set-
tings or patient settings.
Parameter for ventilation or monitoring may
be set user-specific.
After restart of the device, check the settings
and adapt them to your patient, if neccessary.
Summary of Contents for Fabius
Page 36: ...36 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 84: ...84 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 130: ...130 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 192: ...192 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 215: ...Instructions for Use Fabius MRI SW 3 n 215 Technical Data 311 ...
Page 216: ...216 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 218: ...218 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 226: ...226 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 228: ...228 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 231: ...Instructions for Use Fabius MRI SW 3 n 231 This page intentionally left blank ...