Assembly
Connecting the Breathing Hoses and
Breathing Bag
When connecting and disconnecting, always hold
the breathing hoses by the end sleeve, not by the
spiral reinforcement. Otherwise, the spiral rein-
forcement may be torn loose.
Breathing hoses with a damaged spiral reinforce-
ment can kink or become occluded.
Before each use, check the breathing hoses for
damage.
NOTE
Take care not to damage the breathing hoses.
WARNING
Risk of use of toxic or incompatible materials
Breathing hoses used on the Fabius MRI must
comply with current national standards.
WARNING
Do not use antistatic or conductive breathing
hoses or masks. There is a risk of burns in the
MRI environment or when using high fre-
quency electrosurgical equipment.
WARNING
Do not use silicone breathing hoses. They are
visible in MRI images and can be interpreted
as an artifact.
049
Summary of Contents for Fabius
Page 36: ...36 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 84: ...84 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 130: ...130 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 192: ...192 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 215: ...Instructions for Use Fabius MRI SW 3 n 215 Technical Data 311 ...
Page 216: ...216 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 218: ...218 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 226: ...226 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 228: ...228 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 231: ...Instructions for Use Fabius MRI SW 3 n 231 This page intentionally left blank ...