Instructions for Use Fabius MRI SW 3.n
191
Maintenance
z
Remove the absorber canister by turning it
clockwise.
z
Empty the expired CO
2
absorbent from the
absorber into an appropriate refuse container.
z
Fill the absorber with fresh CO
2
absorbent.
1
Fit the absorber canister into position below the
breathing system and turn it counterclockwise
as far as possible.
CLIC Adapter*
The disposable CLIC adapter absorber can also be
used on the Fabius MRI. For information on install-
ing the CLIC adapter, consult its Instructions for
Use.
Checking Readiness for Operation
Perform the Daily and Preuse Checkout procedure
provided in the Appendix of this manual to ensure
that the machine is ready for operation.
NOTE
Ensure that no CO
2
absorbent dust/particles have
been deposited between the gaskets and sealing
surfaces. Such dust and particles can cause leaks
in the system.
*
optional
056
1
Summary of Contents for Fabius
Page 36: ...36 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 84: ...84 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 130: ...130 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 192: ...192 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 215: ...Instructions for Use Fabius MRI SW 3 n 215 Technical Data 311 ...
Page 216: ...216 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 218: ...218 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 226: ...226 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 228: ...228 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 231: ...Instructions for Use Fabius MRI SW 3 n 231 This page intentionally left blank ...