Instructions for Use Fabius MRI SW 3.n
57
Assembly
Connecting Pipeline Supply of N
2
O,
AIR and O
2
1
Connect the gas fitting on each pipeline supply
hose to the corresponding fitting on the rear of
the machine.
2
Connect the other end of each pipeline supply
hose to the appropriate functioning wall outlet.
Pipeline Pressure Gauges
Pipeline pressure gauges for N
2
O (
1
), AIR (
2
), and
O
2
(
3
) are standard. These gauges are located
directly to the right of their corresponding flow con-
trol valves. A typical pressure gauge and flowmeter
assembly is shown in the figure.
WARNING
Carefully check hoses each time you connect
a machine to a wall outlet to ensure that both
ends of the hose are indexed for the same gas.
Pipeline delivery hoses used between wall
outlets and anesthesia machines have caused
accidents when, during assembly, an oxygen
fitting was placed on one end of the hose and
a nitrous oxide fitting on the other end.
CAUTION
To ensure that gas supplies are at adequate pres-
sure, pipeline pressure gauges must indicate
steady pressures of between 41 and 87 psi (2.8
and 6 kPa x 100).
NOTE
The loss of the pipeline supply could cause the
loss of combined devices.
013
1
2
080
1
2
3
Summary of Contents for Fabius
Page 36: ...36 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 84: ...84 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 130: ...130 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 192: ...192 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 215: ...Instructions for Use Fabius MRI SW 3 n 215 Technical Data 311 ...
Page 216: ...216 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 218: ...218 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 226: ...226 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 228: ...228 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 231: ...Instructions for Use Fabius MRI SW 3 n 231 This page intentionally left blank ...