Instructions for Use Fabius MRI SW 3.n
75
Assembly
Auxiliary power outlets
In cases of a blown safety fuse for the auxiliary out-
lets:
– Remedy the fault
– Have the safety fuse replaced by an electrician
WARNING
The auxiliary outlets are not powered by the
uninterruptible power supply (UPS) in the
event of a power failure.
WARNING
Risk of electric shock
Connecting equipment to the auxiliary power
outlets may cause the patient leakage current
to rise above the permitted values if a protec-
tive earth conductor should fail. The risk of
electric shock cannot be excluded in such
cases.
WARNING
Do not connect life-supporting devices to the
auxiliary power outlets of the anesthesia
workstation. The auxiliary power outlets are
not powered by the uninterruptible power sup-
ply (UPS) in the event of a power failure.
WARNING
Risk of increased leakage current
Do not connect HF surgical devices to the
auxiliary power outlets.
WARNING
Risk of increased leakage current
Additional power adapter outlets must not be
connected to the auxiliary power outlets.
Summary of Contents for Fabius
Page 36: ...36 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 84: ...84 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 130: ...130 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 192: ...192 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 215: ...Instructions for Use Fabius MRI SW 3 n 215 Technical Data 311 ...
Page 216: ...216 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 218: ...218 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 226: ...226 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 228: ...228 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 231: ...Instructions for Use Fabius MRI SW 3 n 231 This page intentionally left blank ...