Instructions for Use Fabius MRI SW 3.n
109
Operation
When changing from Volume Control or Pressure
Control to Pressure Support, the suggested value
for
Δ
P
PS
is either the last used value or the site
default value.
When changing from Volume Control or Pressure
Control to Pressure Support, the suggested value
for
Trigger
is either the last used value or the site
default value.
When switching between Volume Control mode
and SIMV/PS mode, the
P
MAX
and
PEEP
settings
shall automatically be transferred from the previous
mode to the new mode.
When switching from Pressure Support mode to
SIMV/PS mode, the
Δ
P
PS
,
Insp Flow
,
Trigger
, and
PEEP
settings shall automatically be transferred
from the previous mode to the new mode.
When switching from SIMV/PS mode with Pressure
Support enabled to Pressure Support mode, the
Δ
P
PS
and
Insp Flow
settings shall automatically be
transferred from SIMV/PS to Pressure Support.
When switching from SIMV/PS mode to Pressure
Support mode, the
Trigger
and
PEEP
settings shall
automatically be transferred from SIMV/PS to Pres-
sure Support.
Summary of Contents for Fabius
Page 36: ...36 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 84: ...84 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 130: ...130 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 192: ...192 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 215: ...Instructions for Use Fabius MRI SW 3 n 215 Technical Data 311 ...
Page 216: ...216 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 218: ...218 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 226: ...226 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 228: ...228 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 231: ...Instructions for Use Fabius MRI SW 3 n 231 This page intentionally left blank ...