Technical Data
196
Instructions for Use Fabius MRI SW 3.n
MRI Use
Usable within magnets of 1.5 tesla and 3 tesla static magnetic field.
Distance to center of magnetic field: fringe field strength of 40 mtesla (400 gauss) or less.
Ambient Conditions
During operation
Temperature
10 to 35 °C (50 to 95 °F)
Atmospheric pressure
700 to 1060 cmH
2
O (hPa)
Relative humidity
20 to 80 % (no condensation)
During storage
Temperature
–10 to 60 °C (14 to 140 °F)
Atmospheric pressure
700 to 1060 cmH
2
O (hPa)
Relative humidity
10 to 90 % (no condensation)
The service conditions for supplementary equipment must be noted. These may restrict the area of use
for the system as a whole. Vaporizer units and the anesthetic agents used may restrict the area of use of
the workstation in regards to its temperature range and maximum fresh-gas flow. The corresponding
Instructions for Use of the supplementary equipment must be noted.
Machine Data
Gas supply from medical gas pipeline system
Pipeline System Pressure Range at Machine Connector
O
2
, N
2
O, AIR:
50 to 55 psi (3.4 to 3.8 kPa x 100)
Note: Pipeline system supply pressure variation
shall not exceed ±10 %
Gas supply connectors:
Each inlet is fitted with a non-return valve
NIST or DISS (where required)
Pipeline Pressure Indicator Accuracy
±3 % of full scale from 40 to 120 psi
(2.7 to 8.3 kPa x 100)
Summary of Contents for Fabius
Page 36: ...36 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 84: ...84 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 130: ...130 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 192: ...192 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 215: ...Instructions for Use Fabius MRI SW 3 n 215 Technical Data 311 ...
Page 216: ...216 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 218: ...218 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 226: ...226 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 228: ...228 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 231: ...Instructions for Use Fabius MRI SW 3 n 231 This page intentionally left blank ...