145
Configuration
Current settings
The current settings cannot be saved as default
settings.
To change the site defaults follow the instructions
on pages “Restore Site Defaults” on page 139 to
page 141.
Restore Factory Defaults
z
Select and confirm the »
Restore Factory
Defaults
« label on the Standby Setup screen.
The Restore Factory Default setting window
appears on the screen.
z
Select and confirm »
Yes
« or »
No
«. When
»
Yes
« is selected, the factory defaults are
restored and replace the current default set-
tings.
The factory defaults for the Fabius MRI are shown
in the table below:
Parameter
Factory Default Settings
Volume
Control
P
MAX
= 40
V
T
= 600
Freq = 12
T
I
:T
E
= 1:2.0
T
IP
:T
I
= 10
PEEP = 0
Pressure
Control
P
INSP
= 15
Freq = 12
T
I
:T
E
= 1:2.0
Insp Flow = 30
PEEP = 0
Pressure
Support
Δ
P
PS
= 10
Freq Min = 3
Trigger = 2
Insp Flow = 30
PEEP = 0
541
Summary of Contents for Fabius
Page 36: ...36 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 84: ...84 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 130: ...130 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 192: ...192 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 215: ...Instructions for Use Fabius MRI SW 3 n 215 Technical Data 311 ...
Page 216: ...216 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 218: ...218 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 226: ...226 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 228: ...228 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 231: ...Instructions for Use Fabius MRI SW 3 n 231 This page intentionally left blank ...