Instructions for Use Fabius MRI SW 3.n
201
Technical Data
Electromagnetic immunity
The Fabius MRI is intended for use in the electro-
magnetic environment specified below. The user of
the Fabius MRI should assure that is used in such
an environment.
Immunity against IEC 60601-1-2 test level Compliance
level (of this
equipment)
Electromagnetic environment
electrostatic dis-
charge, ESD
(IEC 61000-4-2)
contact discharge: ±6 kV ±6 kV
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humid-
ity shall be at least 30 %.
air discharge: ±8 kV
±8 kV
electrical fast tran-
sients / bursts
(IEC 61000-4-4)
power supply lines: ±2 kV ±2 kV
Mains power quality should be that of
a typical commercial or hospital envi-
ronment.
longer input / output lines:
±1 kV
±1 kV
surges on AC
mains lines
(IEC 61000-4-5)
common mode: ±2 kV
±2 kV
Mains power quality should be that of
a typical commercial or hospital envi-
ronment.
differential mode: ±1 kV
±1 kV
power frequency
magnetic field
50/60 Hz
(IEC 61000-4-8)
3 A/m
3 A/m
In close vicinity to the Fabius MRI, no
equipment with extraordinary power
frequency magnetic fields (power
transformers, etc.) should be oper-
ated.
voltage dips and
short interruptions
on AC mains input
lines
(IEC 61000-4-11)
dip >95 %, 0.5 periods
>95 %, 0.5 per. Mains power quality should be that of
a typical commercial or hospital envi-
ronment. If user requires continued
operation during power mains inter-
ruptions, it is recommended to power
the Fabius MRI from an uninterrupt-
ible supply or a battery.
dip 60 %, 5 periods
60 %, 5 per.
dip 30 %, 25 periods
30 %, 25 per.
dip >95 %, 5 seconds
>95 %, 5 sec.
Summary of Contents for Fabius
Page 36: ...36 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 84: ...84 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 130: ...130 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 192: ...192 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 215: ...Instructions for Use Fabius MRI SW 3 n 215 Technical Data 311 ...
Page 216: ...216 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 218: ...218 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 226: ...226 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 228: ...228 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 231: ...Instructions for Use Fabius MRI SW 3 n 231 This page intentionally left blank ...