Operation
100
Instructions for Use Fabius MRI SW 3.n
Pressure Control Ventilation
Starting Pressure Control Ventilation
The following examples and illustrations describe
starting Pressure Control ventilation from the
present ventilation mode “Volume Control”:
1
Press the »
Pressure Control
« key. The LED
associated with this key starts blinking. It
remains blinking until the selected mode of
operation is confirmed.
2
The Waveform window is replaced by the venti-
lator settings window and a message that pro-
vides instructions to confirm the mode change.
z
If the ventilator settings are correct, confirm the
mode change.
z
If the ventilator settings are not correct, for each
parameter that needs to change, press the cor-
responding soft key, select the correct value,
and confirm the change. When the parameter
changes are completed, confirm the ventilation
mode change.
z
After the mode change is confirmed, the
»
Pressure Control
« key LED switches from
blinking to constantly on, the ventilator switches
to the Pressure Control mode, and the wave-
form is restored.
The parameters that can be set for Pressure Con-
trol mode are shown in the adjacent table, along
with their adjustment ranges and factory default
values.
Due to compliance and resistance combinations,
the adjusted Freq Min might not applied accurately
in the Pressure Control Mode.
519
1
2
520
Ventilation Parameter
(Pressure Control
mode)
Adjustment
Range
Factory
Default
Value
Inspiratory Pressure
P
INSP
[cmH
2
O] ([hPa])
5 to 65
min. PEEP+5
15
Frequency
Freq [bpm] ([1/min])
4 to 60
12
Insp time:
Exp time
T
I
:T
E
4:1 to 1:4
1:2
Inspiratory Flow
Insp Flow
[L/min]
10 to 75
30
PEEP [cmH
2
O] ([hPa]) 0 to 20
0
Summary of Contents for Fabius
Page 36: ...36 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 84: ...84 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 130: ...130 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 192: ...192 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 215: ...Instructions for Use Fabius MRI SW 3 n 215 Technical Data 311 ...
Page 216: ...216 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 218: ...218 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 226: ...226 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 228: ...228 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 231: ...Instructions for Use Fabius MRI SW 3 n 231 This page intentionally left blank ...