Assembly
60
Instructions for Use Fabius MRI SW 3.n
1
Open the cylinder valves.
z
To ensure that the cylinder pressures are ade-
quate, check that the cylinder gauges indicate
pressures recommended in the table below.
z
Close the cylinder valves.
Cylinders attached to the hanger yokes must con-
tain gas at the recommended pressures shown in
the table below. (Indicated pressures are for E-size
cylinders at 70 °F/21 °C.) Cylinders measuring less
than the minimum recommended pressure
(PSI - MIN) should be replaced with new, full cylin-
ders.
Dimensions of Usable Cylinders
Gas
PSI/kPa x 100 -
FULL
(typical full load)
PSI/kPa x 100 -
MIN
AIR
1900/131
1000/69
N
2
O
745/51
600/42
O
2
1900/131
1000/69
E-size Pin-Index Cylinder
4.7 L
Height 25.5 in (654 mm)
Diameter 4.3 in (111 mm)
D-size Pin-Index Cylinder (without valve)
4.0 L
Height 25.5 in (654 mm)
Diameter 4.3 in (111 mm)
2.8 L
Height 16.5 in (419 mm)
Diameter 4.3 in (111 mm)
014
1
Summary of Contents for Fabius
Page 36: ...36 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 84: ...84 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 130: ...130 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 192: ...192 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 215: ...Instructions for Use Fabius MRI SW 3 n 215 Technical Data 311 ...
Page 216: ...216 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 218: ...218 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 226: ...226 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 228: ...228 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 231: ...Instructions for Use Fabius MRI SW 3 n 231 This page intentionally left blank ...