Technical Data
210
Instructions for Use Fabius MRI SW 3.n
Breathing System
Volume with reusable absorber
incl. filled absorber, excl. hoses
1.7 L + bag
Volume with Drägersorb CLIC adapter
incl. filled absorber, excl. hoses
1.7 L + bag
excl. absorber, excl. hoses
0.8 L + bag
Volume absorber
reusable absorber, filled
1.5 L
Clic absorber
(Drägersorb CLIC Free)
1.2 L
Compliance
(in automatic ventilation modes, e.g. Volume Control, excluding patient hoses)
with reusable absorber
0.35 mL/cmH
2
O (0.35 mL/hPa)
with Drägersorb CLIC adapter,
including absorber
0.35 mL/cmH
2
O (0.35 mL/hPa)
with Drägersorb CLIC adapter,
excluding absorber
0.35 mL/cmH
2
O (0.35 mL/hPa)
Resistance
(Reusable absorber or CLIC absorber with or without flexible bag arm, normal operation, filled with
Drägersorb 800 +)
Inspiratory
Expiratory
As per ISO 8835-2, dry, max.
±6 hPa (±6 cmH
2
O), with hose
set for adults M30146
–4.4 cmH
2
O
(–4.7 hPa)
4.2 cmH
2
O
(4.4 hPa)
As per ISO 8835-2, dry, sole bre-
athing system without
patient hoses
–3.7 cmH
2
O
(–3.7 hPa)
3.7 cmH
2
O
(3.7 hPa)
Typical leak
<50 mL/min
Summary of Contents for Fabius
Page 36: ...36 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 84: ...84 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 130: ...130 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 192: ...192 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 215: ...Instructions for Use Fabius MRI SW 3 n 215 Technical Data 311 ...
Page 216: ...216 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 218: ...218 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 226: ...226 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 228: ...228 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 231: ...Instructions for Use Fabius MRI SW 3 n 231 This page intentionally left blank ...