Operation
88
Instructions for Use Fabius MRI SW 3.n
Setting Vaporizer Concentration
Refer to the appropriate Instructions for Use for the
vaporizer being used. Vapor 2000 is shown and
described below.
For the Dräger Interlock 2 system:
z
Ensure that the vaporizer is properly seated.
1
Lock the unused vaporizer by moving the selec-
tor lever completely towards it. For example, to
lock the left vaporizer, move the lever to the left.
2
With the handwheel set to »
T
« position on the
unlocked vaporizer, press the button and
engage the handwheel at »
0
«. Wait five sec-
onds for the pressure to balance.
3
Press the button and turn the handwheel coun-
terclockwise to set the required anesthetic gas
concentration.
z
Regularly check the filling level on the sight
glass. When reaching the minimum mark, fill the
vaporizer with anesthetic agent.
For the Dräger Auto Exclusion system:
z
Close any open vaporizers.
z
Ensure that the vaporizer is properly seated.
1
With the handwheel set to »
T
« position, press
the button and engage the handwheel at »
0
«.
Wait five seconds for the pressure to balance.
2
Press the button and turn the handwheel coun-
terclockwise to set the required anesthetic gas
concentration.
z
Regularly check the filling level on the sight
glass. When reaching the minimum mark, fill the
vaporizer with anesthetic agent.
041
1
2
3
072
1
2
Summary of Contents for Fabius
Page 36: ...36 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 84: ...84 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 130: ...130 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 192: ...192 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 215: ...Instructions for Use Fabius MRI SW 3 n 215 Technical Data 311 ...
Page 216: ...216 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 218: ...218 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 226: ...226 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 228: ...228 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 231: ...Instructions for Use Fabius MRI SW 3 n 231 This page intentionally left blank ...