Assembly
50
Instructions for Use Fabius MRI SW 3.n
Activating the Battery
The Fabius MRI anesthesia machine is shipped
with the battery fuse disconnected in order to pre-
vent discharge during shipment and storage prior to
installation.
z
Remove the battery fuse from the top drawer of
the machine.
z
Remove the battery fuse from its packaging.
1
Insert the battery fuse into the battery fuse
holder. Turn the fuse holder 1/4-turn clockwise
until it is snug.
WARNING
For operation in magnetic fields the combina-
tion of vaporizers, anesthesia workstation,
and MRI scanner must be tested by experts
(trained and factory authorized technical ser-
vice representatives for anesthetic machines
and MRI scanners) prior to the first use to
ensure proper vaporizer and interface func-
tion in the specific magnetic fields. Otherwise
uncontrolled concentrations, leakage, and/or
malfunction of the interlock system may
occur. The testing has to take into consider-
ation all positions of the anesthesia worksta-
tion, including the vaporizer, in which it will
operate in the MRI environment during daily
use. Additionally, it is necessary to check if
the imaging of the MRI scanner is adversely
affected by the vaporizer and the anesthesia
workstation.
010
1
Summary of Contents for Fabius
Page 36: ...36 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 84: ...84 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 130: ...130 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 192: ...192 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 215: ...Instructions for Use Fabius MRI SW 3 n 215 Technical Data 311 ...
Page 216: ...216 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 218: ...218 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 226: ...226 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 228: ...228 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 231: ...Instructions for Use Fabius MRI SW 3 n 231 This page intentionally left blank ...