14
Instructions for Use Fabius MRI SW 3.n
For Your Safety and that of Your Patients
CAUTION
Risk of physical injury
To avoid physical injury, pay special attention to
edges, moving parts and corners when working
with
– drawers,
– the ventilator module,
– the writing tray,
– swivel arms for mounted devices,
– gas cylinders,
– vaporizer units,
– CLIC absorbers and CLIC adapters,
as well as other accessories.
CAUTION
Fabius MRI is intended to be used on 120 V /
60 Hz supplies in the USA. The corresponding
power cable is listed in the list of accessories.
Connection of Fabius MRI to a 240 V / 60 Hz sup-
ply is only allowed if it is a center-tapped-single-
phase circuit and the power cable is a certified,
hospital-grade cable 10 A/ 250 V AC of max. 3 m
(9.8 ft) length with:
– a NEMA 6-15 HG plug,
– a SJT 3*18 AWG cable and
– a C13 connector.
In this case Dräger recommends the mains cable
with part no. 1868152.
NOTE
Software must be installed by qualified personnel.
We recommend to contact DrägerService for soft-
ware installation.
NOTE
If the correct functionality of the protective earthing
conductor or its connection to the device is doubt-
ful, the device must be operated using the internal
power supply (battery).
Summary of Contents for Fabius
Page 36: ...36 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 84: ...84 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 130: ...130 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 192: ...192 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 215: ...Instructions for Use Fabius MRI SW 3 n 215 Technical Data 311 ...
Page 216: ...216 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 218: ...218 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 226: ...226 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 228: ...228 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 231: ...Instructions for Use Fabius MRI SW 3 n 231 This page intentionally left blank ...