Instructions for Use Fabius MRI SW 3.n
69
Assembly
Connecting the Pressure Sensor
1
Press the pressure measuring line onto the
hose barb on the underside of the breathing
system until it engages.
Do not squeeze the pressure measuring line
when pressing it onto the hose barb.
2
Connect the pressure measuring line to the
bacterial filter, and plug it firmly onto the port
labeled
on the connector panel on the back
of the machine.
Connecting the Breathing Pressure
Gauge*
z
Push the pressure measuring line onto the hose
barb (
3
), the breathing pressure gauge port (
4
)
to the pressure gauge connector, and then to
the bacterial filter and onto the port labeled
on the connector panel on the back of the
machine (
5
).
*
optional
017
2
1
018
4
3
5
Summary of Contents for Fabius
Page 36: ...36 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 84: ...84 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 130: ...130 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 192: ...192 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 215: ...Instructions for Use Fabius MRI SW 3 n 215 Technical Data 311 ...
Page 216: ...216 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 218: ...218 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 226: ...226 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 228: ...228 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 231: ...Instructions for Use Fabius MRI SW 3 n 231 This page intentionally left blank ...