Instructions for Use Fabius MRI SW 3.n
207
Technical Data
Anesthesia Gas Supply Module
Fresh Gas Flow Indicators:
.
O
2
, N
2
O, AIR:
Range and accuracy: 0.0 to 12.0 L/min ±10 % of reading or
0.12 L/min into an ambient atmosphere of 14.7 psi
(1.013 kPa x 100) at 20 °C (68 °F).
Resolution: 0.1 L/min
Fresh Gas Flow Stability:
O
2
and N
2
O: ±10 % of setting with pipeline pressures between 45 to 65 psi (3.1 to 4.5 kPa x 100)
AIR: ±10 % of setting with pipeline pressures between 50 to 55 psi (3.4 to 3.8 kPa x 100)
Air flow rate will vary proportionally with supply pressures outside 50 to 55 psi (3.4 to 3.8 kPa x 100).
Total Fresh Gas Flowmeter:
Range and accuracy:
0 to 10 L/min ±10 % of full scale at STP,
calibrated with 50 % O
2
/ 50 % N
2
O gas mixture
0 to 10 L/min ±15 % of full scale at STP for all other gas mix-
tures
Resolution:
0.5 L/min from 0.5 to 2 L/min
1.0 L/min from 2 to 10 L/min
O
2
flush (bypass):
at 55 psi (3.8 kPa x 100): max. 50 L/min
at 50 psi (3.4
kPa x 100): min. 35 L/min
Common Gas Outlet Pressure Limit:
13 psi (0.9 kPa x 100), maximum
Auxiliary Oxygen Flowmeter
Connection
Staged connector for use with different hose diameters
Fresh-gas flow
0 to 10 L/min
Accuracy
±5 % of full scale
Resolution
0.5 L/min
Summary of Contents for Fabius
Page 36: ...36 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 84: ...84 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 130: ...130 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 192: ...192 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 215: ...Instructions for Use Fabius MRI SW 3 n 215 Technical Data 311 ...
Page 216: ...216 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 218: ...218 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 226: ...226 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 228: ...228 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 231: ...Instructions for Use Fabius MRI SW 3 n 231 This page intentionally left blank ...