Cleaning
172
Instructions for Use Fabius MRI SW 3.n
Removing the Waste Gas Port
z
Unscrew the waste gas port.
Removing the Flow Sensor
1 L
oosen fitting on the expiration port.
2
Remove the flow-sensor guard.
3
Extract the flow sensor.
CAUTION
Risk of flow measurement failure
Disinfecting or cleaning the flow sensors by
machine will damage them and cause the flow
measurement to fail.
Disinfect and clean the flow sensors as described
in the Instructions for Use of the Spirolog, Infinity
ID flow sensor and SpiroLife flow sensors.
CAUTION
Risk of flow measurement failure
Sterilizing the Spirolog and Infinity ID flow sensors
in high-temperature steam will damage them and
cause the flow measurement to fail.
Disinfect and clean the flow sensor as described
in the Instructions for Use of the Spirolog, Infinity
ID flow sensor and SpiroLife flow sensors.
092
06
3
1
2
3
Summary of Contents for Fabius
Page 36: ...36 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 84: ...84 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 130: ...130 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 192: ...192 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 215: ...Instructions for Use Fabius MRI SW 3 n 215 Technical Data 311 ...
Page 216: ...216 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 218: ...218 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 226: ...226 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 228: ...228 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 231: ...Instructions for Use Fabius MRI SW 3 n 231 This page intentionally left blank ...