Instructions for Use Fabius MRI SW 3.n
177
Cleaning
Other
AGS scavenging
system
Daily
Reprocessing according to appropriate Instructions for Use.
Suction system and
suction bottle
assembly accessories
Daily
Reprocessing according to appropriate Instructions for Use.
1) It is generally recommended to perform a pre-cleaning under running water for approx. 5 minutes.
For dried soiling an additional pre-cleaning step using an ultrasonic bath is recommended. The addition of a quaternary
ammonium compound supports the cleaning performance effectively (e. g. Gigasept AF). Perform pre-cleaning in the ultra-
sonic bath for approx. 15 minutes. Afterwards rinse under clear running water for approx. 5 minutes (waste water must be
free of visible cleaning residues).
2) Use mild alkaline or neutral cleaner
3) Use disinfectants based on aldehydes and quaternary ammonium compounds.
4) Do not use any agents containing alcohol.
5) After the machine cleaning and disinfection, a hot steam sterilization is required to dry the breathing system. Insufficient
drying of the control areas found in the valve plate can lead to an adverse effect on the functions of the medical device or
to its failure!
6) Valves, ventilator cover and AGS sleeve: Make sure that the cleaning and rinsing liquids can flow in the opening direction
of the valve.
7) Remove any water which may have accumulated in the ventilator diaphragm. Large quantities of condensed water can
negatively effect the operation of the device and lead to failure of the medical device!
Fabius MRI
components which
can be reprocessed
Recommen-
ded
reprocessing
intervals
Manual pre-
cleaning
1)
Machine
cleaning and
thermal
disinfection
2)
Manual disinfection
3)
Hot steam
sterili-
zation
Surface
disinfection
Immersio
n bath
disinfec-
tion
Summary of Contents for Fabius
Page 36: ...36 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 84: ...84 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 130: ...130 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 192: ...192 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 215: ...Instructions for Use Fabius MRI SW 3 n 215 Technical Data 311 ...
Page 216: ...216 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 218: ...218 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 226: ...226 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 228: ...228 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 231: ...Instructions for Use Fabius MRI SW 3 n 231 This page intentionally left blank ...