Configuration
140
Instructions for Use Fabius MRI SW 3.n
Standby Setup Screen
Pressing the Setup key while in Standby provides
access to various default and configuration set-
tings. The settings made in this screen will be
saved as Site Defaults.
In order for these settings to go into effect, press
the »
Restore Site Defaults
« key on the Standby
screen. The Site Defaults are also restored any
time power is cycled or the system test is performed
(see page 133).
1
Press the »
Setup
« key while in Standby mode.
The operator is requested to enter a four-digit pass-
word in order to prevent unauthorized changes to
the basic functions. This password is allocated
when commissioning the workstation. If desired,
DrägerService can set an individual password or
disable this functionality.
z
Select and confirm the figures successively
from the line displayed using the rotary knob.
z
The Standby Setup screen replaces the
Standby screen. The cursor allows the user to
select »
Default Settings
« (see below) or
»
Configuration
« (see page 147). (Selecting
and confirming the return arrow on the right of
the Setup screen will exit the Standby Setup
screen and redisplay the Standby screen.)
554
1
504
Summary of Contents for Fabius
Page 36: ...36 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 84: ...84 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 130: ...130 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 192: ...192 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 215: ...Instructions for Use Fabius MRI SW 3 n 215 Technical Data 311 ...
Page 216: ...216 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 218: ...218 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 226: ...226 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 228: ...228 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 231: ...Instructions for Use Fabius MRI SW 3 n 231 This page intentionally left blank ...