Assembly
68
Instructions for Use Fabius MRI SW 3.n
Inserting a new O
2
Sensor Capsule
1
Unscrew the cap from the sensor housing.
z
Remove the new sensor capsule from its pack-
aging.
2
Insert the capsule in the housing, with the ring-
shaped conductors against the contacts in the
housing.
z
Screw the cap on firmly by hand.
Connecting the O
2
Sensor
3
Push the O
2
sensor into the port opening of the
inspiratory port dome, and
4
Plug the connector into the fitting labeled
on the connector panel on the back of the
machine.
051
1
2
016
3
4