40
Instructions for Use Fabius MRI SW 3.n
Operation Concept
Rotary Knob
1
The rotary knob is the main control used to
select and confirm all monitoring and system
settings.
–
Turn
the rotary knob to change or select a value
or parameter (clockwise rotation increases a
value; counterclockwise rotation decreases a
value). This function is indicated in the exam-
ples and instructions in this manual by “select.”
–
Press
the rotary knob to set a value or confirm
a selection. If the selection is not confirmed, the
value or parameter will not change. This func-
tion is indicated in the examples and instruc-
tions in this manual by “confirm.”
Fixed Function Keys
The fixed function keys positioned on both sides of
the screen provide access to major machine and
monitoring functions. Most fixed function keys must
be confirmed by pressing the rotary knob.
2
The »
ManSpont
« key selects the Man-
ual/Spontaneous ventilation mode.
3
The »
SIMV/PS
« key is used to select the
SIMV/PS ventilation mode (optional).
4
The »
Pressure Support
« key is used to select
the Pressure Support ventilation mode
(optional).
5
The »
Pressure Control
« key is used to select
the Pressure Control ventilation mode.
6
The »
Volume Control
« key is used to select
the Volume Control ventilation mode.
7
The »
Alarms
« key displays the Alarm Limits
window.
8
The »
Setup
« key provides two different func-
tions, depending on mode:
– If pressed during Standby mode, it displays the
Standby Setup screen which enables the user
to define site defaults and configure system set-
tings (see page 132).
– If pressed during a ventilation mode, it allows
the user to view and change monitoring settings
9
The »
Home
« key displays the main screen from
any other screen currently displayed
10
The »
Silence
« key silences all active alarm
tones for two minutes.
11
The »
Standby
« key switches the machine to
standby mode. Monitoring and alarms are
turned off and the ventilator stops. Fresh gas
monitoring continues.
566
1
11
10
2
3
4
5
6
7
8
9
Summary of Contents for Fabius
Page 36: ...36 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 84: ...84 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 130: ...130 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 192: ...192 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 215: ...Instructions for Use Fabius MRI SW 3 n 215 Technical Data 311 ...
Page 216: ...216 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 218: ...218 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 226: ...226 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 228: ...228 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 231: ...Instructions for Use Fabius MRI SW 3 n 231 This page intentionally left blank ...