Instructions for Use Fabius MRI SW 3.n
181
Cleaning
Reinstalling the Scavenger System
Reconnecting the Anesthetic Gas Receiving
System (AGS)
Every Anesthetic Gas Scavenging System (AGSS)
used on the Fabius MRI must follow ISO standard
8835-3.
The scavenging system is used with vacuum
waste-gas disposal systems.
The AGS or passive scavenger is not applied as an
independent system.
It is used as one of the three components of the
AGSS.
z
Hook receiving system with the slots over the
appropriate pins of the basic device and allow it
to slide down into place.
1
Connect the scavenging hose to the relevant
socket of the receiving system.
z
Connect the connector of the scavenging hose
to the terminal unit of the disposal system.
2
Close the connection that is not used with a
screw cap.
z
Push the transfer hose on the designated
socket.
z
Connect the other end of the transfer hose to
the waste-gas port located on the underside of
the breathing system.
WARNING
Risk of patient injury
If the side openings of the receiving system
are blocked, negative pressure may result in
the breathing system and the patient’s lungs.
Always make sure the side openings of the
receiving system are not blocked.
NOTE
Remove the socket from the scavenger hose
before connecting.
NOTE
The scavenger hoses must not be pinched,
kinked, or blocked in any manner.
061
2
1
Summary of Contents for Fabius
Page 36: ...36 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 84: ...84 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 130: ...130 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 192: ...192 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 215: ...Instructions for Use Fabius MRI SW 3 n 215 Technical Data 311 ...
Page 216: ...216 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 218: ...218 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 226: ...226 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 228: ...228 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 231: ...Instructions for Use Fabius MRI SW 3 n 231 This page intentionally left blank ...