Operation
112
Instructions for Use Fabius MRI SW 3.n
Ending Operation
1
Press and confirm the »
Standby
« key.
Monitoring and alarms are turned off and the venti-
lator stops.
2
Turn vaporizers off by turning the handwheel to
»
0
« until the button engages.
z
Turn off fresh gas flow on the Fabius MRI.
Sleep mode will activate 2.5 minutes after fresh
gas is turned off.
z
Close the cylinder valves.