Instructions for Use Fabius MRI SW 3.n
135
Configuration
Calibrate O
2
Sensor
z
Press the »
Calibrate O2 Sensor
« key on the
Standby screen. The Standby soft key window
is replaced by instructions for performing the
calibration.
z
Follow the instructions on the screen:
– Remove O
2
sensor and expose to room air for
2 minutes.
– To start O
2
Calibration press rotary knob.
z
When the calibration begins, the instructions
are removed, and a »
O
2
Calibration in
progress
« message is displayed above the
Standby soft keys.
z
When the calibration is completed, one of two
messages is posted above the Standby soft
keys: »
O
2
Sensor Calibration completed -
reinsert O
2
sensor
« or »
O
2
Sensor Calibra-
tion Failed
«.
Troubleshooting O
2
Sensor Calibration Failure
If the O
2
sensor cannot be calibrated, replace the
O
2
capsule in the O
2
sensor housing (see
page 189).
If the O
2
sensor still cannot be calibrated, call
DrägerService or your local authorized service
organization.
598
529
530
531
532
Summary of Contents for Fabius
Page 36: ...36 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 84: ...84 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 130: ...130 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 192: ...192 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 215: ...Instructions for Use Fabius MRI SW 3 n 215 Technical Data 311 ...
Page 216: ...216 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 218: ...218 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 226: ...226 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 228: ...228 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 231: ...Instructions for Use Fabius MRI SW 3 n 231 This page intentionally left blank ...