Instructions for Use Fabius MRI SW 3.n
101
Operation
Based on a defined frequency (
Freq
) and a defined
ratio of inspiratory time to expiratory time (
Ti:TE
), a
volume is applied. This volume depends on the set
inspiratory pressure (
P
INSP
) and the patient compli-
ance. The
Insp Flow
key is used to set the slope of
the pressure waveform. In addition, a positive end-
expiratory pressure (
PEEP
) can be set.
The lower pressure limit
P
AW
low
is used for pres-
sure monitoring to detect apneas (disconnection)
and continuous pressure. An alarm is generated
when the pressure curve does not cross the pres-
sure threshold either from above or below.
11
0
1
P
AW
low
PEEP
T
I
T
E
Flow
P
INSP
P
AW
Time
Time
Freq
Summary of Contents for Fabius
Page 36: ...36 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 84: ...84 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 130: ...130 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 192: ...192 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 215: ...Instructions for Use Fabius MRI SW 3 n 215 Technical Data 311 ...
Page 216: ...216 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 218: ...218 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 226: ...226 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 228: ...228 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 231: ...Instructions for Use Fabius MRI SW 3 n 231 This page intentionally left blank ...