Fault-Cause-Remedy
Priority Alarm Message
Probable Cause
Remedy
!!
(20)
APNEA VENTILATION
Breathing/ventilation stops.
Breathing circuit disconnected
or leakage.
If two or more consecutive
Apnea Ventilation breaths are
auto triggered.
Pressure Support settings are
incorrect.
Check ventilator.
Check breathing circuit.
A spontaneous patient breath
is detected by the Fabius MRI.
Check Pressure Support set-
tings.
!
(7)
BATTERY LOW
AC failure and battery <20 %
Restore mains power.
!!
(17)
BATTERY LOW
AC failure and battery <10 %
Restore mains power.
!!!
(26)
CHECK APL VALVE
APL bypass valve fault.
Check ventilator diaphragm
and close cover.
Check APL bypass valve con-
nection for disconnect or leak.
Select Standby Mode and
switch back to the previous
ventilation mode.
Check the APL valve setting.
!
(7)
CHECK BATTERY
The reserve power is 0 % of a
full charge.
Replace fuse.
Call DrägerService or your
local authorized service orga-
nization.
!!!
(31)
CONTINUOUS PRESSURE
Breathing pressure above
threshold for more than
15 seconds.
Check breathing circuit. If in
ManSpont mode, check fresh
gas flow.
Check actual limit
for
"Pressure treshold alarm
limit".
!!
(15)
EXP PORT LEAKAGE
Expiratory flow of more than
15 mL measured during
inspiration in Volume Control,
Pressure Control, or Pressure
Support mode.
Check expiratory valve and
valve disk.
Check tubing of expiration
control line.
Check flow sensor.
Follow the procedure to cali-
brate flow sensor.
Call DrägerService or your
local authorized service orga-
nization.
!!
(16)
EXP PRESSURE HIGH
PEEP is more than 4 cmH
2
O
(hPa) above the PEEP setting
in an automatic ventilation
mode.
Check PEEP/P
MAX
, etc. hoses
for kinks.
Summary of Contents for Fabius
Page 36: ...36 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 84: ...84 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 130: ...130 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 192: ...192 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 215: ...Instructions for Use Fabius MRI SW 3 n 215 Technical Data 311 ...
Page 216: ...216 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 218: ...218 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 226: ...226 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 228: ...228 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 231: ...Instructions for Use Fabius MRI SW 3 n 231 This page intentionally left blank ...