Instructions for Use Fabius MRI SW 3.n
171
Cleaning
Removing the Compact Breathing Sys-
tem
z
Remove all breathing hoses.
z
Disconnect the breathing bag extension and
bag by loosening the two thumb screws.
z
Remove the ventilation hose.
z
Remove the fresh gas hose from the breathing
system.
z
Remove the scavenger hose.
z
Remove the flow sensor cable.
z
Remove the O
2
sensor cable.
z
Remove the breathing pressure cable.
z
Detach the APL-bypass and the PEEP/P
MAX
lines from the breathing system and from the
side of the machine.
z
Remove the absorber (for complete instruc-
tions, see page 189
z
Remove the compact breathing system.
1
Remove plug for the inspiration dome.
Removing the Inspiratory Valve
2
Unscrew the retaining nut.
3
Remove the inspection cap.
4
Extract the valve disc.
5
Extract the gasket on top of the valve disc.
Removing the Expiratory Valve
6
Unscrew the retaining nut.
7
Remove the inspection cap.
8
Extract the valve disc.
9
Remove the gasket on top of the valve disc.
060
2
3
4
6
7
8
1
5
9
Summary of Contents for Fabius
Page 36: ...36 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 84: ...84 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 130: ...130 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 192: ...192 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 215: ...Instructions for Use Fabius MRI SW 3 n 215 Technical Data 311 ...
Page 216: ...216 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 218: ...218 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 226: ...226 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 228: ...228 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 231: ...Instructions for Use Fabius MRI SW 3 n 231 This page intentionally left blank ...