Instructions for Use Fabius MRI SW 3.n
137
Configuration
Compliance Test Results
The system compliance test determines the current
compliance of the patient system with any filters,
hoses, and Y-piece. The compliance value is used
during volume controlled ventilation for applying a
tidal volume matching accurately the set tidal vol-
ume. A system compliance value of up
6.5 mL/cmH
2
O (6.5 mL/hPa) is posted as
»
PASSED
« on the leak test result screen, and the
compliance value is also displayed on the Standby
screen.
Ventilator Leak Test Results
If a ventilator leak test is posted as »
PASSED
«, the
value is displayed on the leak test result screen and
the Standby screen.
System Leak Test Results
System leak test results are posted on the leak test
result screen.
Measured ventilator
leak [mL/min]
Displayed result
[mL/min]
≤
150
measured value and
PASSED
151 to 250
measured value and
FAILED
>250
>250 and FAILED
Measured system leak
[mL/min]
Displayed result
[mL/min]
≤
250
measured value and
PASSED
251 to 350
measured value and
FAILED
>350
>350 and FAILED
11
6
O
2+
Ventilator
Fresh-Gas
Decoupling
Vaporizer
Flow Tube
Oxygen
Flush
Sample Line: closed
Absorber
APL
Bypass
APL
Valve
Setting
P
MAX
/
PEEP
Bag
Scavenging
Y-piece: closed
11
5
O
2+
Ventilator
Fresh-Gas
Decoupling
Vaporizer
Flow Tube
Oxygen
Flush
Sample Line: closed
Absorber
APL
Bypass
APL
Valve
Setting
P
MAX
/
PEEP
Bag
Scavenging
Y-piece: closed
Summary of Contents for Fabius
Page 36: ...36 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 84: ...84 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 130: ...130 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 192: ...192 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 215: ...Instructions for Use Fabius MRI SW 3 n 215 Technical Data 311 ...
Page 216: ...216 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 218: ...218 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 226: ...226 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 228: ...228 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 231: ...Instructions for Use Fabius MRI SW 3 n 231 This page intentionally left blank ...