Technical Data
208
Instructions for Use Fabius MRI SW 3.n
Anesthetic Agent Vaporizer Interface
Dräger Halothane Vapor 2000
Dräger Enflurane Vapor 2000
Dräger Isoflurane Vapor 2000
Dräger Sevoflurane Vapor 2000
Dräger Vapor quick-change plug-in system for up to two anesthetic agent vaporizers.
The connections are automatically closed and sealed when the vaporizer is removed.
Vapor 2000 system only
See specific Instructions for Use manuals for technical data of anesthetic agent vaporizers.
Monitoring and Measurement Display Range
Resolution
Accuracy
Condition
P
AW
Airway pressure (numeric)
–20 to 99 cmH
2
O
(hPa)
1 cmH
2
O
(hPa)
±4 %
1)
1) Max. ±4 % of the measured value or ±2 cmH
2
O (±2 hPa), whichever is greater.
Airway pressure (wave)
0 to 99 cmH
2
O
(hPa)
Pressure gauge (mechanical)
–20 to 80 cmH
2
O
(hPa)
5 cmH
2
O
(hPa)
1.6 cmH
2
O
(hPa)
V
E
Expiratory minute volume
0 to 99.9 L/ min 0.1 L/ min
±15 %
2)
2) At standard test conditions per IEC 60601-2-13.
with reference
to 68 °F (20 °C),
ambient pres-
sure and satu-
rated gas
Expiratory tidal volume
0 to 1500 mL
1 mL
±15 %
2)
or
±20 mL,
whichever is
greater
f
Breathing frequency
2 to 99 bpm
(1/min)
±1 bpm
(1/min)
±1 bpm
(1/min)
FiO
2
O
2
measurement in the
main gas flow
10 to 100 vol.% 1 vol.%
±(2.5 vol.%
+ 2.5% of the
measured
value) as per
ISO 21647
with reference
to ambient pres-
sure during cali-
bration
Summary of Contents for Fabius
Page 36: ...36 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 84: ...84 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 130: ...130 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 192: ...192 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 215: ...Instructions for Use Fabius MRI SW 3 n 215 Technical Data 311 ...
Page 216: ...216 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 218: ...218 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 226: ...226 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 228: ...228 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 231: ...Instructions for Use Fabius MRI SW 3 n 231 This page intentionally left blank ...