Assembly
58
Instructions for Use Fabius MRI SW 3.n
Connecting the Reserve Gas Cylinders
for N
2
O, AIR and O
2
( for Pin-Index
Mounting)
WARNING
Use only non-magnetic (e.g., aluminum) cylin-
ders with the Fabius MRI. Steel cylinders will
cause serious injury or death if brought into
an MR scanning room.
WARNING
When attaching a cylinder, ensure that only
one washer is installed between the cylinder
and the yoke gas inlet. The use of multiple
washers will inhibit the pin-index safety sys-
tem. Be sure to verify the presence of the
index pins each time a cylinder is installed.
Never attempt to override the pin-index safety
system.
WARNING
Do not oil or grease the O
2
cylinder valves and
O
2
pressure regulator. There is a risk of explo-
sion.
CAUTION
Even if the gas supply is connected to a medical
gas pipeline, the cylinders should remain on the
device in reserve.
NOTE
If cylinder valves are leaky or difficult to open or
close, they must be repaired in accordance with
the manufacturer’s specifications.
Summary of Contents for Fabius
Page 36: ...36 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 84: ...84 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 130: ...130 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 192: ...192 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 215: ...Instructions for Use Fabius MRI SW 3 n 215 Technical Data 311 ...
Page 216: ...216 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 218: ...218 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 226: ...226 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 228: ...228 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 231: ...Instructions for Use Fabius MRI SW 3 n 231 This page intentionally left blank ...