Instructions for Use Fabius MRI SW 3.n
103
Operation
Starting Pressure Support Ventilation
The following examples and illustrations describe
starting Pressure Support ventilation from the
present ventilation mode “Volume Control”:
1
Press the »
Pressure Support
« key. The LED
associated with this key starts blinking. It
remains blinking until the selected mode of
operation is confirmed.
2
The Waveform window is replaced by the venti-
lator settings window and a message that pro-
vides instructions to confirm the mode change.
z
If the ventilator settings are correct, confirm the
mode change.
z
If the ventilator settings are not correct, for each
parameter that needs to change, press the cor-
responding soft key, select the correct value,
and confirm the change. When the parameter
changes are completed, confirm the ventilation
mode change.
z
After the mode change is confirmed, the
»
Pressure Support
« key LED switches from
blinking to constantly on, the ventilator switches
to the Pressure Support mode, and the wave-
form is restored.
564
2
1
52
1
Summary of Contents for Fabius
Page 36: ...36 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 84: ...84 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 130: ...130 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 192: ...192 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 215: ...Instructions for Use Fabius MRI SW 3 n 215 Technical Data 311 ...
Page 216: ...216 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 218: ...218 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 226: ...226 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 228: ...228 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 231: ...Instructions for Use Fabius MRI SW 3 n 231 This page intentionally left blank ...