Instructions for Use Fabius MRI SW 3.n
139
Configuration
Access Alarm Log
1
Press the »
Access Alarm Log
« key on the
Standby screen. The Standby soft key window
is replaced by the alarm log, which lists all
alarms with their dates and times.
z
To scroll through the alarm log, turn the rotary
knob.
z
To delete all alarms from the log, select and
confirm the »
Clear Alarm Log
« label.
z
To exit the alarm log and return to the Standby
screen, select and confirm the return arrow.
Restore Site Defaults
2
Press the »
Restore Site Defaults
« key on the
Standby screen. The predefined site defaults
are restored, and a »
Site Default settings
restored
« message is displayed above the
Standby soft keys.
The site default settings are password protected.
Changes of the settings can be made via the
Standby Setup screen, see page 140.
CAUTION
Alarm log data is deleted if Fabius MRI is swit-
ched off or a total loss of electrical power supply
occurs.
WARNING
Risk of inappropriate ventilation settings
After site default settings have been restored
check whether the ventilation and monitoring
settings are appropriate to the patient.
56
1
53
4
561
535
1
2
Summary of Contents for Fabius
Page 36: ...36 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 84: ...84 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 130: ...130 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 192: ...192 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 215: ...Instructions for Use Fabius MRI SW 3 n 215 Technical Data 311 ...
Page 216: ...216 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 218: ...218 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 226: ...226 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 228: ...228 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 231: ...Instructions for Use Fabius MRI SW 3 n 231 This page intentionally left blank ...