Technical Data
204
Instructions for Use Fabius MRI SW 3.n
Electrical Safety Conformance
Conforms to:
z
UL 60601-1
z
IEC 60601-1
z
CAN/CSA C22.2 No. 601.1-M90
General Safety Standards for
Anesthesia
z
IEC 60601-2-13 plus US deviations
z
ISO 8835-2
z
EN 740
Freedom from latex
Fabius MRI is latex-free!
Latex-free breathing bags and breathing hoses
must be used for latex-free use.
Ventilator
Conforms to:
z
ISO 8835-5
Control Inputs Ranges
P
MAX
Pressure limiting
15 to 70 cmH
2
O (1 cmH
2
O resolution)
(15 to 70 hPa (1 hPa resolution))
(setting must be at least 10 cmH
2
O (10 hPa) above
PEEP; and in SIMV/PS mode, the PMAX setting must
also be greater than
Δ
P
PS
+PEEP)
V
T
Tidal volume
20 to 1400 mL (10 mL resolution)
V
T
(SIMV/PS)
Tidal volume
20 to 1100 mL (10 mL resolution)
f
Breathing frequency
4 to 60 bpm (1/min) (1 bpm resolution)
(4 to 60 1/min (1/min resolution))
T
I
/T
E
Inspiration/expiration ratio
4 : 1 to 1 : 4
T
IP
/T
I
Inspiration pause
0 % to 50 % (1 % resolution)
PEEP
End-expiratory pressure
0 to 20 cmH
2
O (1 cmH
2
O resolution)
(0 to 20 hPa (1 hPa resolution))
P
INSP
Inspiratory pressure
5 to 65 cmH
2
O (1 cmH
2
O resolution)
(5 to 65 hPa (1 hPa resolution))
(setting must be at least 5 cmH
2
O (5 hPa) above
PEEP)
Summary of Contents for Fabius
Page 36: ...36 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 84: ...84 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 130: ...130 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 192: ...192 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 215: ...Instructions for Use Fabius MRI SW 3 n 215 Technical Data 311 ...
Page 216: ...216 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 218: ...218 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 226: ...226 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 228: ...228 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 231: ...Instructions for Use Fabius MRI SW 3 n 231 This page intentionally left blank ...