176
Instructions for Use Fabius MRI SW 3.n
Cleaning
Care List for Fabius MRI Components
The following table lists Fabius MRI components
with recommended processing methods. Process-
ing refers to cleaning, disinfection, and/or steriliza-
tion, as appropriate for a given component. The
table is intended as a guide. Follow the institution’s
policies regarding specific methods and agents for
cleaning and sterilization.
Applicable for non-infectious patients
The list is only intended as a rough guideline. The
instructions of the hospital´s hygiene officer shall
prevail and must be observed!
CAUTION
Fabius MRI and its components must not be trea-
ted with formaldehyde vapors or ethylene oxide.
CAUTION
For infectious patients, all parts that come into
contact with breathing gas also have to be sterili-
zed after disinfection and cleaning.
Fabius MRI
components which
can be reprocessed
Recommen-
ded
reprocessing
intervals
Manual pre-
cleaning
1)
Machine
cleaning and
thermal
disinfection
2)
Manual disinfection
3)
Hot steam
sterili-
zation
Surface
disinfection
Immersio
n bath
disinfec-
tion
Surfaces
Device surfaces
Per patient
No
No
Yes
4)
No
No
Power supply cable,
pressure gas hoses
Per patient
No
No
Yes
4)
No
No
Respiratory pressure
meter
Daily
No
No
Yes
No
No
Components for carrying breathing gas
Breathing system
Per patient
Yes
Yes
5)
No
Yes
Yes
Inspiratory valve,
Expiratory valve, APL
valve
Per patient
Yes
6)
Yes
No
Yes
Yes
Expiratory port
Per patient
Yes
Yes
No
Yes
Yes
Exhaust port valve
Per patient
Yes
6)
Yes
No
Yes
Yes
Ventilator cover
Per patient
Yes
6)
Yes
No
Yes
Yes
Ventilator diaphragm
7)
Per patient
Yes
Yes
No
Yes
Yes
Ventilator hose
Per patient
No
Yes
5)
No
Yes
Yes
Absorber and insert
Per patient
Yes
Yes
No
Yes
Yes
Infinity ID/Spirolog/
SpiroLife flow sensor
Note the Instructions for Use of the flow sensors
Manual bag valve
mask holder
Per patient
Yes
Yes
No
Yes
Yes
Summary of Contents for Fabius
Page 36: ...36 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 84: ...84 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 130: ...130 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 192: ...192 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 215: ...Instructions for Use Fabius MRI SW 3 n 215 Technical Data 311 ...
Page 216: ...216 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 218: ...218 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 226: ...226 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 228: ...228 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 231: ...Instructions for Use Fabius MRI SW 3 n 231 This page intentionally left blank ...