Instructions for Use Fabius MRI SW 3.n
63
Assembly
Connecting the Passive Scavenger*
The passive scavenger system is used only with
non-recirculating exhaust systems. It is not meant
to be used with vacuum disposal systems.
z
Install the passive scavenger on the machine by
sliding its bracket onto the two shoulder screws
located on the scavenger mounting bracket on
the side of the machine.
1
Connect one end of the transfer hose to the side
fitting on the scavenger.
z
Connect the other end of the transfer hose to
the waste gas port located on the underside of
the breathing system.
2
Connect the waste gas hose to the bottom con-
nection on the scavenger.
z
Connect the other end of the hose to the hospi-
tal waste gas disposal system.
For detailed information on the passive scavenger
system, refer to separate Instructions for Use.
*
optional
NOTE
The scavenger hoses must not be pinched,
kinked, or blocked in any manner.
NOTE
Remove the socket from the scavenger hose
before connecting.
062
1
2
Summary of Contents for Fabius
Page 36: ...36 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 84: ...84 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 130: ...130 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 192: ...192 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 215: ...Instructions for Use Fabius MRI SW 3 n 215 Technical Data 311 ...
Page 216: ...216 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 218: ...218 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 226: ...226 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 228: ...228 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 231: ...Instructions for Use Fabius MRI SW 3 n 231 This page intentionally left blank ...