Breathing Pressure Monitoring Alarms
While the ventilator is on, apnea pressure alarms
are generated if the breathing pressure monitor
does not sense a valid breath for a specified period
of time (see »
APNEA PRESSURE
« on page 162).
While the ventilator is off and the system is in Man-
Spont mode, these alarms are generated at
30 seconds (Caution) and 60 seconds (Warning).
Setting Pressure High and Threshold Alarm
Limits
Follow the procedure "Setting Alarm Limits" on
page 121 to change the breathing pressure high or
threshold alarm limit.
NOTE
The pressure threshold alarm limit should be as
close as possible to the sensed plateau pressure
without exceeding it, approximately 4 cmH
2
O
(hPa) below the plateau pressure.
Summary of Contents for Fabius
Page 36: ...36 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 84: ...84 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 130: ...130 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 192: ...192 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 215: ...Instructions for Use Fabius MRI SW 3 n 215 Technical Data 311 ...
Page 216: ...216 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 218: ...218 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 226: ...226 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 228: ...228 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 231: ...Instructions for Use Fabius MRI SW 3 n 231 This page intentionally left blank ...