Monitoring
128
Instructions for Use Fabius MRI SW 3.n
Breathing Pressure Monitoring
Breathing Pressure Monitoring Windows
The breathing pressure monitoring windows show
the following breathing pressure information in
numeric and graphic form:
1 PEEP
(Positive End Expiratory Pressure)
shows the breathing pressure at the end of
exhalation in cmH
2
O (hPa). The display range
is from 0 to 30 cmH
2
O (0 to 30 hPa).
2 PLAT
(Plateau) Breathing Pressure shows the
breathing pressure at the end of inspiration in
cmH
2
O (hPa). The display range is from
0 to 80 cmH
2
O (0 to 80 hPa)
or
2 MEAN
Breathing Pressure shows the average
of all the instantaneous pressure values
recorded during each breath in cmH
2
O (hPa).
The display range is from 0 to 50 cmH
2
O
(0 to 50 hPa).
3 PEAK
Breathing Pressure shows the highest
instantaneous pressure value for each breath in
cmH
2
O (hPa). The display range is from
0 to 80 cmH
2
O (0 to 80 hPa).
4
Pressure High Alarm Limit
5
Pressure Threshold Alarm Limit
6
Breathing Pressure Trace Window displays a
breathing pressure trace (waveform).
7
Breathing Pressure Threshold Limit Line
8
Breathing Pressure Minimum and Maximum
Trace Scale Limits Indicator. Pressure mea-
surements are automatically scaled from
0 to 20, 0 to 50, or 0 to 100 cmH
2
O
(0 to 20, 0 to 50, or 0 to 100 hPa).
NOTE
The Fabius MRI can be configured by
DrägerService or your local authorized service
organization to display mean pressure (MEAN)
instead of plateau pressure (PLAT).
513
1
2
3
6
7
8
4
5
Summary of Contents for Fabius
Page 36: ...36 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 84: ...84 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 130: ...130 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 192: ...192 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 215: ...Instructions for Use Fabius MRI SW 3 n 215 Technical Data 311 ...
Page 216: ...216 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 218: ...218 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 226: ...226 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 228: ...228 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 231: ...Instructions for Use Fabius MRI SW 3 n 231 This page intentionally left blank ...