Configuration
136
Instructions for Use Fabius MRI SW 3.n
Leak/Compliance Test
The »
Leak/Compl Test
« key is used to initiate a
system compliance test, a system leak test, and a
ventilator leak test.
1
Press the »
Leak/Compl Test
« key on the
Standby screen. The Standby soft key window
is replaced by a ventilator preparation message
and then instructions for performing the test.
z
Follow the instructions on the screen:
– Close all fresh gas control valves.
– Place Y-piece on plug located on bag mount.
– Seal sample-line connector (if applicable).
– Switch APL valve to “MAN” and set pressure to
40 to 50.
– Press O
2
flush button, build up a pressure of
15 to 30.
z
Upon completion of the test, the results are
posted on the screen.
z
Press the rotary knob to return to the Standby
screen.
56
1
597
533
59
6
1
Summary of Contents for Fabius
Page 36: ...36 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 84: ...84 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 130: ...130 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 192: ...192 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 215: ...Instructions for Use Fabius MRI SW 3 n 215 Technical Data 311 ...
Page 216: ...216 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 218: ...218 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 226: ...226 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 228: ...228 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 231: ...Instructions for Use Fabius MRI SW 3 n 231 This page intentionally left blank ...