Configuration
154
Instructions for Use Fabius MRI SW 3.n
Calibrate O
2
Sensor
To calibrate the oxygen sensor correctly, make sure
it is exposed only to room air during the entire cali-
bration period.
To avoid leakage, remove the oxygen sensor
assembly from the inspiratory valve dome, and
insert the valve dome plug into the dome.
1
Press the »
Calibrate O2 Sensor
« soft key in
the Setup window.
z
The calibration instruction window replaces the
Setup window. Follow the directions provided:
– Remove O
2
Sensor and expose to room air for
2 minutes.
– To start O
2
Calibration press rotary knob.
– Observe Calibration status in O
2
data window.
– Reinsert O
2
Sensor after successful Calibra-
tion.
503
1
514
Summary of Contents for Fabius
Page 36: ...36 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 84: ...84 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 130: ...130 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 192: ...192 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 215: ...Instructions for Use Fabius MRI SW 3 n 215 Technical Data 311 ...
Page 216: ...216 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 218: ...218 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 226: ...226 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 228: ...228 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 231: ...Instructions for Use Fabius MRI SW 3 n 231 This page intentionally left blank ...